Vehicle report

2020 DEMERS AMBULANCES MXP153 DODGE RAM

Safety recalls

1 campaign

21V740000

STRUCTURE:BODY:ROOF AND PILLARS Recall

Braun Ambulances is recalling certain 2020 Demers Ambulances MXP153 Ford F-Series Gas, MXP153 Dodge Ram, and 2020-2021 MXP153 Ford F-Series Diesel emergency vehicles. The medical oxygen tube and electrical wires may become damaged due to insufficient clearance from the ceiling spacers.

Remedy: Dealers will remove the ceiling panels and spacers, replace wires and oxygen network tubing, and create a routing channel inside the insulation board for all wires, free of charge. Owner letters were mailed on October 6, 2021. Owners may contact Braun Ambulances' customer service at 1-877-344-9990.

Reading this report

Complaint counts are reports filed with NHTSA, not failure rates. A common model collects more of them than a rare one that breaks more often.